12/26/2004
0:58 TRD latch-up (cal trigger is 0.6Hz) HP
1:00 TRD started working after first ped-run file HP
1:15 TRD latch-up (cal trigger is 0.2Hz) HP
1:21 TRD started working after first ped-run file HP
1:26 TRD latch-up (cal trigger is 2.0Hz) HP
1:31 TRD started working after first ped-run file HP
1:48 TRD latch-up (cal trigger is 2.0Hz) HP
1:51 TRD started working after first ped-run file HP
2:10 TRD latch-up (cal trigger is 0.1Hz) HP
2:12 TRD started working after first ped-run file HP
2:41 TRD latch-up (cal trigger is 1.2Hz) HP
2:52 TRD started working after the second ped-run file HP
2:54 Send "get_spi.stol" YS
3:16 TRD latch-up (cal trigger is 1.2Hz) HP
3:22 TRD started working after first ped-run file HP
3:33 TRD latch-up (cal trigger is 3.0Hz) Hp
3:41 Cal trigger rate become around 10-18 YS
Buffer (PKT0 goes up to 5%)
Next command window will start 4:24(EST)
3:50 Hoskeeping does not updtaed. Data flow looked stopped YS
ESC said they are going to make it back.
4:02 ESC said data is still coming from sfc, but there is some problem on ground computer. They are going to reboot their computer  YS
4:07 Connrction comes back.. YS
Cal trigger rate is still too high about 10-18..
4:14 Buffer (PKT0) became 20%. Command window will be open in 10 minutes YS
4:19 TRD starts working after  the fourth ped-run file HP
4:22 TRD latch-up (cal trigger is 17.0Hz) HP
4:25 cal trigger threshold 70 YS
CAL trigger rate become around 1-3 Hz YS
Buffer(PKT0) goes down slowly. YS
4:56 TRD starts working after  the fourth ped-run file HP
4:58 TRD latch-up (cal trigger is 2.0Hz) HP
5:20   Buffer(PKT0) is cleared 0% YS
5:23 TRD started working after the third ped-run file HP
5:32 TRD latch-up (cal trigger is 3.0Hz) HP
5:33 TRD started working after the first ped-run file HP
5:40 TRD latch-up (cal trigger is 1.0Hz) HP
5:43 TRD started working after the first ped-run file HP
5:50 TRD latch-up (cal trigger is 2.6Hz) HP
5:53 TRD started working after the first ped-run file HP
6:04 TRD latch-up (cal trigger is 2.6Hz) HP
6:14 TRD started working after the first ped-run file HP
6:21 TRD latch-up (cal trigger is 4.3Hz) HP
6:23 TRD started working after the first ped-run file HP
6:29 TRD latch-up (cal trigger is 2.6Hz) HP
6:33 TRD started working after the first ped-run file HP
6:39 TRD latch-up (cal trigger is 2.6Hz) HP
6:43 TRD started working after the first ped-run file HP
6:49 TRD latch-up (cal trigger is 5.0Hz) HP
6:54 TRD started working after the first ped-run file HP
7.09 TRD latch-up (cal trigger is 5.0Hz) HP
7:12 Send "get_spi.stol" YS
7:24 TRD started working after the first ped-run file HP
7:26 TRD latch-up (cal trigger is 6.0Hz) HP
7:34 TRD started working after the first ped-run file HP
7:39 TRD latch-up (cal trigger is 6.0Hz) HP
7:44 TRD started working after the first ped-run file HP
7:59 TRD latch-up (cal trigger is 2.3Hz) HP
8:04 TRD started working after the first ped-run file HP
8:19 TRD latch-up (cal trigger is 2.5Hz) HP
8:34 TRD started working after the second ped-run file HP
8:40 TRD latch-up (cal trigger is 7.9Hz) HP
9:03 TRD started working after the third ped-run file HP
9:07 TRD latch-up (cal trigger is 2.0Hz) HP
9:23 TRD started working after the second ped-run file HP
9:29 TRD latch-up (cal trigger is 1.4Hz) HP
9:46 TRD started working after data file HP
9:54 TRD latch-up (cal trigger is 3.2Hz) HP
9:56 TRD started working after data file HP
10:01 TRD latch-up (cal trigger is 3.4Hz) HP
10:09 cal trigger threshold 72 JH
10:17 TRD started working after data file HP
10:19 TRD latch-up (cal trigger is 1.4Hz) HP
10:26 TRD started working after data file HP
10:41 TRD latch-up (cal trigger is 1.2Hz) HP
10:46 TRD started working after data file HP
11:04 TRD latch-up (cal trigger is 0.6Hz) HP
11:17 TRD started working after data file HP
11:09 TRD latch-up (cal trigger is 0.0Hz) HP
11:16 TRD started working after data file HP
11:28 ssssssss HP
11:47 TRD started working after data file HP
11:51 TRD latch-up (cal trigger is 1.4Hz) HP
12:07 TRD started working after data file HP
12:24 TRD latch-up (cal trigger is 0.8Hz) HP
12:27 TRD started working after data file HP
13:25 changed pedestal run every 5 minutes JH
13:45 TRD latch-up (cal trigger is 1.2Hz) HP
13:46 TRD started working after data file HP
14:29 TRD latch-up (cal trigger is 1.2Hz) HP
14:37 TRD started working after data file HP
14:54 TRD latch-up (cal trigger is 0.2Hz) HP
14:57 TRD started working after data file HP
15:26 TRD latch-up (cal trigger is 0.5Hz) HP
15:27 TRD started working after data file HP
15:29 Send "get_spi.stol" JH
15:30 TRD latch-up (cal trigger is 0.5Hz) HP
15:32 TRD started working after data file HP
15:42 TRD latch-up (cal trigger is 1.0Hz) HP
15:47 TRD started working after data file HP
17:23 TRD latch-up (cal trigger is 0.4Hz) HP
17:24 TRD started working after data file HP
18:03 TRD latch-up (cal trigger is 0.2Hz) HP
18:08 TRD started working after data file HP
18:12 rebalance top TRD JH
ramp down TRD HV1, open valve 32-35, close valve 32-35
ramp up TRD HV1 to 1.53KV
rebalance bottom TRD
ramp down TRD HV2, open valve 10-13, close valve 10-13
ramp up TRD HV2 to 1.56KV
the above two are required by Simon
19:35 TRD latch-up (cal trigger is 0.2Hz) HQ
19:40 TRD started working after data file HQ
20:10 TRD latch-up (cal trigger is 0.2Hz) HQ
20:15 TRD started working after the first ped-run file HQ
20:35 TRD latch-up (cal trigger is 0.6Hz) JH
20:40 TRD started working after opening data file JH
21:35 TRD latch-up (cal trigger is 0.2Hz) JH
21:41 TRD started working after the first ped-run file HQ
22:06 Send "get_spi.stol" JH
23:52 TRD latch-up (cal trigger is 0.2Hz) HQ
23:57 TRD started working after the first ped-run file HQ