| Remember trigger threshold resets to default value whenever you send "SCI_SPARE1 0" |
| powering off command box causes resetting cal boxes to lose HV |
| Must ramp down HV & bias before turning off SFC |
| Consider having ESC reset SFC before taking SFC power procedure |
| See operation manual for the Emergency TRD shut down procedure that Simon provided 12/23/04 |
| If TRD should be off/on, set TRD HV as following |
| HV 1 (TRD1) 1870V |
| HV 5 (TRD2) 1900 V |
| HV 9 (Cherenkov) 1225 V |
| Check HSK monitoring values should be as following |
| HV1 = 1.77 KV |
| HV2 = 1.82 KV |
| HV3 = 1.18 KV |
| If TCD should be off/on check if level 3 trigger is 300 mV. This is different from the default 350 mV. |
| CAL threshold conversion: DAC/3*2.5 MeV, e.g., DAC70 = 58MeV |